What are the installation and maintenance requirements for the ASP Station 2000 RPS20B cooling unit?

After mounting, let the unit rest for at least 12 hours before powering it on to protect the cooling system. The compartment temperature is adjustable between +2 and +20 °C (default is +4 °C). The condenser should be regularly cleaned with compressed air to remove dust.

V Z

Linda V.

What are the installation and maintenance requirements for the ASP Station 2000 RPS20B cooling unit?

After mounting, allow the unit to rest for 12 hours before powering it on. The sample compartment temperature is adjustable between +2 and +20 °C (default is +4 °C). The condenser, located in the upper part of the device, should be regularly cleaned with compressed air.