How to remove potential air bubbles from the ÄKTA pure system?
To eliminate air bubbles, prime the inlets and purge the pump heads. Fill all inlet tubings with the appropriate buffer, ensuring all inlet tubings to be used are immersed in the correct buffer. Purge the system pumps to remove air from the pump heads by connecting a syringe to the purge valve of each pump head and slowly drawing liquid until it reaches the pump, then close the valve and disconnect the syringe. Ensure no column is online during these operations. Regularly check the flow restrictor and clean the check valves if necessary, as contaminant build-up can also affect the correct functioning of the system.