User questions

A X

Michael A.

What are the methods for setting up measurements with the Spraytec system?

The two main methods are using a Standard Operating Procedure (SOP) via File New SOP... for predefined templates, and manual measurement to configure parameters on the fly or for testing. An optimized manual measurement can be saved as an SOP by selecting Extract SOP.

A X

Michael A.

How to configure measurements using default SOPs or in manual mode?

For default SOPs, select File New SOP... and choose the appropriate template. For manual measurement, go to Measure Manual, then click the Settings button in the Measurement Manager to define your parameters. An optimized manual configuration can be saved as an SOP by right-clicking the record and selecting Extract SOP.

A X

Michael A.

How to determine the necessary parameters in the Measurement Manager?

To determine the necessary parameters, start by using the default SOP templates. The Inspection tab allows real-time observation of the obscuration histogram to verify settings. Ensure the correct lens (300mm or 750mm) is selected based on the expected particle size. For a new product, use a manual measurement as a “temporary SOP” to test settings before saving the configuration.

A X

Michael A.

How are measurements configured using the Spraytec system, including parameter determination for aerosols?

To configure measurements with the Spraytec system and determine necessary parameters:

1. General Configuration Methods:

  • Standard Operating Procedure (SOP): This is the recommended method. An SOP acts as a template defining all measurement parameters and settings. Create or modify an SOP via the software by selecting File–New–SOP... and following the wizard. The software provides default SOP models for continuous sprays, quick measurements, and nasal sprays.
  • Manual Measurement: This allows configuration on the fly, just before measurement. It's ideal for testing new settings (“temporary SOP”) or when parameters for a new spray are uncertain. Access via Measure–Manual.

2. Specific Considerations for Aerosols:

  • Preparation: Shake the container and spray for a few seconds to clean the nozzle before measurement.
  • Safety: Use an extraction system to direct the spray through the measurement zone and protect personnel.
  • Lens Selection: Ensure the correct lens (300mm or 750mm) is selected, adapted to the expected particle size range.

3. Determining Parameters in Measurement Manager:

  • From Measure–Manual, click the Settings button in the Measurement Manager.
  • Start with Defaults: Utilize the default SOP models as a baseline for recommended values.
  • Real-time Inspection: The Inspection tab (available upon launch) allows observing the obscuration and light energy histogram in real-time. Use this to confirm correct settings (positioning, concentration) before validating the measurement.
  • Iterative Testing: For new products, use manual measurement as a “temporary SOP” to test the impact of settings.
  • Saving Custom SOPs: Once parameters are optimized, save a manual measurement configuration as an SOP by right-clicking on the record in the Measurement file window and selecting Extract SOP.
A X

Michael A.

How to configure measurements with the Spraytec system?

Les mesures peuvent être configurées en utilisant deux méthodes principales : les Procédures Opératoires Standard (S.O.P) ou le mode de mesure manuel. Les S.O.P sont des modèles recommandés qui définissent tous les paramètres et réglages de mesure, créables via File–New–SOP.... Le mode manuel permet de configurer les paramètres à la volée, idéal pour tester de nouveaux réglages, et peut être enregistré comme S.O.P par la suite.

A X

Michael A.

How to determine the necessary measurement parameters in the Measurement Manager?

Pour déterminer les paramètres nécessaires, utilisez les modèles de S.O.P par défaut, l'onglet Inspection pour un contrôle en temps réel des réglages, adaptez la lentille (300mm ou 750mm) à la taille des particules attendues, et effectuez des tests itératifs via des mesures manuelles pour les nouveaux produits.

A X

Michael A.

How to configure measurements with the Spraytec system?

Measurements with the Spraytec system can be configured using a Standard Operating Procedure (SOP) as a template by selecting File–New–SOP..., or by using manual measurement to configure parameters on the fly, which can then be saved as an SOP.

A X

Michael A.

How to configure measurements using default SOPs or in manual mode within the Measurement Manager?

To use default SOPs, select File–New–SOP... and choose a suitable model. For manual configuration, access Measure–Manual, then in the Measurement Manager, click the Settings button to define parameters. After optimization, a manual configuration can be saved as an SOP by right-clicking the recording in the Measurement file window and selecting Extract SOP.

A X

Michael A.

How to determine the necessary parameters within the Measurement Manager?

To determine the necessary parameters, start with the default SOP models. Use the Inspection tab during launch to observe the obscuration and light energy histogram, ensuring settings (positioning, concentration) are correct. Select the appropriate lens (300mm or 750mm) based on expected particle size. For new products, use manual measurement as a temporary SOP to test settings before saving.