How does the TitroLine 5000 determine the equivalent volume for titration in automatic mode?
In automatic titration mode, the TitroLine 5000 offers different calculation methods to determine the equivalent volume: The system can evaluate the equivalence point (EQ) of the titration curve to determine the equivalent volume. For endpoint titration (EP), the consumption at the endpoint is used as the result. For pH and mV, it is possible to perform a two-endpoint titration.
J N

Mary J.

How does the TitroLine 5000 determine the equivalence point (EQ)?
The equivalence point (EQ) is determined from the maximum of the 1st derivative of the measurement data.